Inogen One G3 Replacement Columns
Inogen One G3 Replacement Columns

Inogen One G3 Replacement Columns

Brand/Manufacturer: INOGEN
Save

Inogen One G3 Replacement Columns is replacement sieve beds for the Inogen G3 oxygen concentrator. They are easy to replace and the replacement process can be completed in minutes. They should be replaced in every 12 to 18 months to maximize the life of the G3 unit. The LCD screen will display “O2 Service Soon” for a low oxygen purity alert and when the columns need to be changed. To remove the columns, simply press the latch release button located at the bottom of the unit then slide the columns (metal tubes) out of the unit.

Item #DescPkgRewardPrice 
RP-320 For Inogen One G3 with 1-4 Setting 2/Pack $4.95
$165.99
$99.00
RP-321 For Inogen One G3 with 1-5 Setting 2/Pack $4.95
$192.75
$99.00

More Information

Inogen One G3 Column Change Procedure:

  1. Turn off the Inogen One G3 concentrator by pressing the power button to shut down the device.
  2. Remove the Inogen One G3 concentrator from the carry bag.
  3. Remove the battery from the Inogen One G3 concentrator.
  4. Turn Inogen One G3 over so that is standing in the upside down position where the display panel is facing down and the bottom of the Inogen One G3 is facing up.
  5. There are two columns (metal tubes) on the Inogen One G3 concentrator and one is located on each side of the device.
  6. Remove column by pressing the latch button with thumb or finger.
  7. While holding the button open, slide the column (metal tube) out by gripping the exposed aluminum surface with other hand.
  8. Remove the column (metal tube) completely from the Inogen One G3.
  9. Repeat steps 6-8 to remove the other column (metal tube).
  10. Remove dust caps of new column (metal tube). Make sure there is no dust or debris where the dust caps were located.
  11. Insert column (metal tube) into the Inogen One G3 concentrator. Do not leave the column exposed; it should be inserted into the Inogen One G3 as soon as the dust caps have been removed.
  12. Push the column (metal tube) into the device such that the column (metal tube) is fully seated into the Inogen One G3 concentrator. The spring loaded latch button should fully return to the closed position.
  13. Repeat steps 10-12 to install the other column (metal tube).
  14. Connect the AC power supply cord to the Inogen One G3 and plug the power supply AC cord into an electrical outlet. Do Not Power on the Inogen One G3 concentrator.
  15. Press and hold the light button for 10 seconds and screen will display “sieve reset” message. Release button once message is displayed on screen.
  16. Press the bell button once and screen will display “sieve reset successful”.
  17. Press the power button to turn on the Inogen One G3, and use normally.