Inogen One G5 Replacement Columns

Inogen One G5 Replacement Columns

Brand/Manufacturer: INOGEN
Hpfy Stores Save

Inogen One G5 Replacement Columns are removable metal columns that provide a quick and easy way to service the Inogen One G5. They contain the sieve beds that convert normal air with lower levels of oxygen concentration to supply a high concentration of oxygen. These sieve beds wear down after prolonged use.

Why choose Inogen One G5 Replaceable Sieve Beds?

  • Can be easily removed and replaced by customers
  • No tools required to switch out the column pair
  • No need to send the product for repair
  • Save significant time and money of customers and providers
  • Sold and used as a pair

Features of Inogen One G5 Replacement Column Pair

  • Hassle-free Sieve Beds - Customers have the freedom to order and replace sieve beds themselves, thereby saving overhead and shipping costs.
  • Less Inventory - By providing replaceable sieve beds, HME providers don’t need to keep loaner POCs on hand for customers to use during maintenance.
  • Reduced Shipping Costs - We limit the number of deliveries required to replace a single sieve bed while also shipping filters in smaller packages.
  • Save on Maintenance Costs - Allowing the customer to easily replace their own sieve beds saves you the time and money required to perform maintenance/replacements.

When to replace Inogen One G5 Sieve Beds?

Inogen recommends replacing the columns every 18 months to two years. In certain operating conditions, earlier replacement may be required. The Inogen One G5 will let you know when it is time for replacement. For optimal sieve bed (columns) life, the product should be used frequently.
 

Traditional vs. Inogen Filter Replacements

Traditional Filter Replacements Inogen Filter Replacements
  1. Customer contacts the provider to request new filter
  2. Provider sends out a loaned POC for customer to use while their filter is being replaced
  3. Customer ships their POC to the provider for filter replacement
  4. Provider performs maintenance and replaces the sieve beds
  5. Provider ships the customer’s POC back to them
  6. Customer ships loaned POC back to provider
  1. Customer contacts the provider to request new filter
  2. Provider ships the sieve bed to the customer (small package)
  3. Customer receives sieve bed and replaces it themselves within minutes

 

How to replace Inogen One G5 Columns?

Follow these step-by-step instructions to easily switch metal columns without using tools and without having to send the Inogen One G5 in for repair like you would with other portable oxygen concentrators:
How to change Inogen One G5 Replacement Columns?

  1. Turn off the Inogen One G5 concentrator by pressing the power button to shut down the device.
  2. Remove the Inogen One G5 concentrator from the carrying case (if attached).
  3. Remove the battery from the Inogen One G5 concentrator.
  4. Place the Inogen One G5 concentrator on its side so that the underside is visible. The metal column assembly can be seen on one side of the device.
  5. Unlock the column assembly by pushing the latch button away from the columns.
  6. While holding the button open, slide the column assembly out of the device by pulling on the column pull handle.
  7. Remove the columns completely from the Inogen One G5. Both columns are removed as one piece.
  8. Remove dust caps of new column assembly. Make sure there is no dust or debris where the dust caps were located.
  9. Insert column assembly into the Inogen One G5 concentrator. Do not leave the column ends exposed; column assembly should be inserted into the Inogen One G5 as soon as the dust caps have been removed.
  10. Push the column assembly into the device such that the columns are fully seated into the Inogen One G5 concentrator. The spring loaded latch button should fully return to the closed position.
  11. Connect the AC power supply cord to the Inogen One G5 and plug the power supply AC cord into an electrical outlet. Do Not power on the Inogen One G5 concentrator.
  12. Press and hold the plus (+) and (-) minus buttons for 5 seconds. The screen will display "sieve reset" message. Release buttons once message is displayed on screen.
  13. Press the alert/bell button once and screen will display "sieve reset success". Press the power button to turn on the Inogen One G5, and use normally.
  14. If you are using Inogen Connect App, navigate to the Advanced screen, then to Additional Information screen and click on the Column Reset button.
    Note: Column change instructions are only to be used when maintenance is required and are not intended for practice purpose.


Inogen One G5 Column Change Procedure
 