Get 5% Off, Code: SNSIA, Exclusions Apply
Advanced Search
O2 Concepts Oxlife Independence Portable Oxygen Concentrator takes POC performance to the next level. It boasts the longest battery life in class and delivers over three times the flow of the leading POC with patented EnergySmart technology. With the Oxlife Independence by users side, they enjoy unlimited oxygen. With DNA technology, it is the world first smart POC. DNA means you can have the machine alert a caregiver or loved one if there is ever an issue with your ambulatory oxygen.
Note: Prescription can be emailed or faxed to 1.203.616.2851. To email please: Click Here
6 Payments: Starts From $333.33
Oxlife Independence User Manual
Device Setup
Device Dimensions with Handle/Wheels
Height: 20.29 in H: 51.5 cm
Width: 10.85 in W: 27.55 cm
Depth: 9.45 in D: 24 cm
WEIGHT
Device Weight
16.7 lbs. (7.57 kg)
Cart Kit (Wheels & Pull Handle)
2.17 lbs. (0.98 kg)
Battery
1.4 lbs. (0.63 kg)
AC Charger
1.9 lbs. (0.86 kg)
DC Charger
.37 lbs. (0.16 kg)
MODES OF OPERATION
Continuous Flow
0.5 to 3 LPM in 0.5 LPM increments; Measured flow values are within ± 2% of actual values. Max flow 3.0 LPM. Flow maintained with outlet pressure ranging from 0 to 5.0 psig (0-34kPa)
Pulse Dose
0.5 to 6.0 (8mL-96mL) setting increments; Measured flow values are within ± 2% of actual values
Battery Specification
14.4V Lithium Ion Battery
External Power Supply Power Input
AC power: 100 - 240 VAC, 50/60 Hz
@ 2.5 amps
DC power: 12-15 VDC; Recommended: 15A outlet at 12V
Altitude
0-13,123ft or (0-4000m)
Oxygen Purity
91% ± 4% (87 - 95%) from 0.5 to 3.0 LPM; Measured purity values are within ± 2% of actual values; There is no variation in purity within the operating altitude.
Operating Temperature
50°F (10°) to 104°F (40°C)
Operating Humidity
10% - 95% @ 82.4°F (28°C)
Operating Atmospheric Pressure
101kPa to 63kPa
Operating Environment
Free of smoke, pollutants, and fumes.
Transport/ Storage Temperature
-4°F (-20°C) to 140°F (60°C)
Transport/ Storage Humidity
0 -95% non-condensing
Operating Time
24 hours per day when connected to an external AC or DC power source. This is a continuous operation device.
Cannula Specification
DO NOT use cannula tubing longer than 7ft (2.13m) when using Pulse Flow Mode
Do NOT use cannula tubing longer than 50ft (15.25m) when using Continuous Flow Mode.
A-Weighted Sound Pressure Level
The volume that the device will reach at maximum settings will be ~ 56dBA with maximum peaks of 58.8dBA.
Alarm Sound Level
All alarms triggered by the device will be 85dBA at 10cm from the unit.
Maximum Outlet Pressure
Device will maintain flow up to 5psi back pressure. Maximum back pressure is 7.0psi.
Limited 5 Year Warranty on Portable Oxygen Concentrator, 1 Year on Accessories Click Here for Warranty
{{x.ShortDescription}} ...