Particulars
|
Specifications
|
Weight
|
7.4oz (210grams) (with alkaline batteries)
|
Oxygen Saturation Display Range
|
0 to 100% SpO2
|
Pulse Rate Display Range
|
18 to 321 beats per minute (BPM)
|
Saturation Declared Accuracy (70 to 100%):
No Motion
|
Adults/Pediatrics:
Finger Clip Sensors- ± 2 digits
Flex Sensors- ± 3 digits
|
Saturation Declared Accuracy (70 to 100%):
Motion
|
Adults/Pediatrics:
Finger Clip and Flex Sensors- ± 3 digits
|
Saturation Declared Accuracy (70 to 100%):
Low Perfusion
|
Adults/Pediatrics:
Finger Clip, Flex and Flexi-Form II Sensors- ± 3 digits
Neonates:
Flex and Flexi-Form II Sensors- ± 4 digits
|
Pulse Rate Declared Accuracy:
No Motion (18 to 300 BPM)
|
Adults/Pediatrics:
Finger Clip, Flex and Flexi-Form II Sensors- ± 3 digits
Neonates:
Flex and Flexi-Form II Sensors- ± 3 digits
|
Pulse Rate Declared Accuracy:
Motion (40 to 240 BPM)
|
Adults/Pediatrics:
Finger Clip Sensors- ± 5 digits
Flex Sensors- ± 3 digits
|
Pulse Rate Declared Accuracy:
Low Perfusion (40 to 240 BPM)
|
Adults/Pediatrics:
Finger Clip, Flex and Flexi-Form II Sensors- ± 3 digits
Neonates:
Flex and Flexi-Form II Sensors- ± 3 digits
|
Measurement Wavelengths and Output Power
|
Red: 660 nanometers @ 0.8mW maximum average
Infrared: 910 nanometers @ 1.2mW maximum average
|
Indicators
|
Pulse Quality Display: LED, tricolor
Numeric Displays: 3-digit 7-segment LEDs, red
Low Battery Indicator: LED, amber
|
Temperature
|
Operating:
-20 to +50°C (-4 to +122°F)
Storage/Transportation:
-30 to +50°C (-22 to +122°F)
|
Humidity
|
Operating: 10 to 90% noncondensing
Storage/Transportation: 10 to 95% noncondensing
|
Altitude
|
Operating: Up to 12,000meters (40,000 ft)
Hyperbaric Pressure: Up to 4 atmospheres
|
Power Requirements
|
(4) 1.5V AA-size alkaline batteries (80 hours typical operation)
|